Determination of 5 Volatile Organic Compounds in Coatings by Headspace Gas Chromatography-Mass Spectrometry
  
View Full Text    Download reader
DOI:10.3969/j.issn.1004-4957.年份.月份
KeyWord:headspace gas chromatography-mass spectrometry(HS/GC-MS)  volatile organic compound  coatings
  
AuthorInstitution
L Qing, ZHANG Qing, KANG Suyuan, BAI Hua, WANG Chao 中国检验检疫科学研究院工业品检验研究所
Hits: 2437
Download times: 913
Abstract:
      A headspace(HS) coupled with gas chromatography-mass spectrometric(GC-MS) method was developed for the determination of 5 residual volatile organic compounds(VOCs)in coating samples. The experimental conditions, such as solvent type, equilibrium temperature and time, as well as the analytical conditions of GC-MS, were optimized. The samples were first dissolved in water-N,N dimethylformamide (1∶1, by volume)and treated at 90 ℃ for 90 min,then separated on a DB VRX capillary column(30 m×025 mm×14 μm),and determined by GC-MS with the external standard method. The results indicated that the recoveries were in the range of 82%-105% with relative standard deviations(RSDs) of 05%-39%. The linear range of 5 VOCs were in the range of 0013 3-667 mg/L with correlation coefficient no less than 0998 9. The limits of quantitation (S/N=10) were no more than 10 mg/kg. The method was used to determine 5 VOCs in 22 kinds of solvent based coating and 31 kinds of waterborne coating samples. The result indicated that the proposed method was simple, accurate, rapid and sensitive and could be applied in the rapid analysis of residual VOCs in various coatings.
Close