Preparation and Characterization of a Mercury(Ⅱ) selective Polymeric Membrane Electrode
  
View Full Text    Download reader
DOI:10.3969/j.issn.1004-4957.年份.月份
KeyWord:ion selective electrode  mercury ion(Ⅱ)  Schiff base
  
AuthorInstitution
ZHANG Jun1,2,ZHAO Yueqian1,DING Jiawang1,QIN Wei1 1、中国科学院烟台海岸带研究所山东省海岸带环境过程重点实验室;2、海南医学院热带医学与检验医学院
Hits: 2019
Download times: 976
Abstract:
      A new Schiff base compound derived from 1,2,4 triazole was synthesized with a high yield.Potentiometric responses of poly(vinyl) chloride(PVC) electrode based on this ligand as neutral ionophore showed high sensitivity and selectivity towards Hg2+ ion.Effects of different plasticizers and amounts of ion exchanger on the potentiometric response property of Hg2+ selective electrode(Hg2+-ISE) were discussed.The result indicated that,under the optimal membrane composition [PVC:3270,o NPOE:6539,ionphore:099,ion exchanger:092,(w∶w)],a good Nernstian response slope of (293±03) mV/dec was obtained for Hg2+ in a wide concentration range of 10×10-6-30×10-4 mol/L with a detection limit of 26×10-7 mol/L Except for silver ion,most of alkali metal,alkaline earth metal and other common transition metal ions didnt interfere the performance of the membrane electrode.The fabricated Hg2+-ISE can be used in a wide pH range of 28-56 with a fast response time of less than 12 s and can be used for 3 month without any obvious change in potentiometric response.The proposed electrode was also successfully applied as indicator electrode in the potentiometric detection of Hg2+ in lab water by standard addition method and the results were confirmed by ICP-MS.