Analysis of Carbonyl Compounds in Water-based Adhesive by Direct Derivatization and High-performance Liquid Chromatography
  
View Full Text    Download reader
DOI:10.3969/j.issn.1004-4957.年份.月份
KeyWord:direct derivatization  solvent extraction  high-performance liquid chromatography  carbonyl compounds  water based adhesive
  
AuthorInstitution
万小红,吴名剑,蒋新宇,戴云辉,李绍晔 1.中南大学化学化工学院;2.湖南中烟工业有限责任公司
Hits: 2661
Download times: 1078
Abstract:
      A direct derivatization and high-performance liquid chromatographic method was developed for the trace analysis of 8 carbonyl compounds,e.g.formaldehyde,acetaldehyde,acetone,acrolein,propaldehyde,crotonaldehyde and 2-butanone and butaldehyde in water-based adhesive.The method adopted 2,4-dinitrophenylhydrazine(DNPH) to directly react with the adhesive sample solution,and the influence factors were investigated and optimized.The best derivatization condition was using the adhesive to react directly with DNPH at 40 ℃ for 20 min.The optimized chromatograpic conditions were as follows:DIONEX Acclaim Explosives E2 reversed-phase column as separation column,water and acetonitrile as mobile phase by double gradient elution.The flow rate was set at 1.2 mL/min,and the column temperature was set at 35 ℃.Under the optimal conditions,the linear ranges of 8 carbonyl compounds were all in 0.001 1-55.32 mg/L,with correlation coefficients higher than 0.999.The limits of detection for 8 carbonyl compounds were 0.066 3-1.691 1 mg/kg,and the limits of quantitation were 0.221-5.637 mg/kg.The relative standard deviations were between 4.3% and 9.3%,and the recoveries were 62%-100%.The established method was sensitive,simple and accurate,and was practical for the determination of 8 carbonyl compounds in wate-based adhesive.
Close