Electrochemical Behavior and Determination of Uric Acid at Reduced Graphene Oxide Modified Electrode
  
View Full Text    Download reader
DOI:10.3969/j.issn.1004-4957.年份.月份
KeyWord:graphene  modified electrode  uric acid  electrochemical behavior
  
AuthorInstitution
朱伟琼,李容,熊健,苟兴龙 西华师范大学化学化工学院化学合成与污染控制重点实验室
Hits: 2979
Download times: 1105
Abstract:
      Electrochemical behavior and response of uric acid(UA) was investigated at the chemically reduced graphene oxide(RGO)modified glassy carbon(GC) electrode(RGO/GC) by cyclic voltammetry,chronocoulometry and AC impedance method.The results showed that,in pH 60 PBS buffer solution,the electrochemical oxidation of UA at the RGO/GC modified electrode was irreversible and controlled by diffusion.Some kinetic parameters such as the number(n) of the electron transferred,the effective area(A) of the modified electrode and the diffusion coefficient(D) were calculated to be 2,0.182 cm2 and 1.51×10-6 cm2?s-1,respectively.A good linear relationship between peak current and concentration of UA was observed in the range of 5.0×10-6-1.5×10-4 mol/L,which met the equation of Ip(μA) =6.457+367.5c(mmol/L)(r=0.995 7).The RGO/GCE modified electrode was successfully applied in the determination of UA with a detection limit of 2.7×10-7 mol/L and recoveries of 98%-100% .
Close