Determination of Phenolic Endocrine Disrupting Chemicals in Human Urine by High Performance Liquid Chromatography Tandem Electrospray Ionization Mass Spectrometry
  
View Full Text    Download reader
DOI:10.3969/j.issn.1004-4957.年份.月份
KeyWord:HPLC-MS/MS  endocrine disrupting chemicals(EDCs)  urine
  
AuthorInstitution
夏同伟,刘良坡,张文静,申河清 1.中国地质大学(武汉)环境学院;2.中国科学院城市环境研究所;3.厦门大学海洋与地球学院
Hits: 2501
Download times: 1369
Abstract:
      A high performance liquid chromatography tandem electrospray ionization mass spectrometry(HPLC-ESI MS/MS) method was developed and validated for the determination of phenolic endocrine disrupting chemicals(EDCs),including bisphenol A(BPA),triclosan(TCS),17α-ethynylestradiol(EE2),estrone(E1) and 17 β-estradiol(E2)in human urine.The samples were purified by liquid-liquid extraction,derivertized with dansyl chloride,and detected by HPLC-ESI MS/MS.The target compounds were quantified by the stable isotope dilution technique.The calibration curve were linear in the range of 0.2-100 μg/L for the 5 EDCs with correlation coefficients more than 0.99.The limits of detection of 5 EDCs were in the range of 0.02-0.27 μg/L.The matrix recoveries of the method for 5 EDCs at three spiked levels of 5,10 and 50 μg/L ranged from 85% to 125%.The relative standard deviations(RSDs,n=3) were between 0.53% and 14.4%.The method was successfully applied in the analysis of 5 EDCs in human urine with its sensitivity and accurancy .