Preparation of A Sensor Based on Graphene Modified Electrode and Molecularly Imprinted Polymers and Its Determination on Dopamine |
|
View Full Text Download reader |
DOI:10.3969/j.issn.1004-4957.年份.月份 |
KeyWord:graphene dopamine molecular imprinted polymers(MIPs) electrochemical sensor |
Author | Institution |
王丹丹,曾延波,刘海清,尹争志,李蕾 |
1.常州大学化学与化工学院;2.嘉兴学院生物与化学工程学院 |
|
Hits: 3522 |
Download times: 1659 |
Abstract: |
An electrochemical sensor based on dropped molecular imprinted polymers(MIPs) suspension for dopamine was fabricated.Influences of pH value,concentrations of graphene and MIPs were also optimized.The optimal conditions were as follows:pH value:7.0,concentration of graphene:0.5 g/L,concentration of MIPs:20 g/L.Under the optimal conditions,the prepared sensor showed good selectivity and high sensitivity for dopamine.The DPV peak current is linear to the DA concentration in the range of 2.0×10-7-1.0×10-4 mol/L,with a detection limit(S/N=3) of 6.8×10-8 mol/L.The prepared sensor was successfully applied in the determination of dopamine hydrochloride with recoveries of 98%-105%. |
Close |