Analysis of Nonyl Phenol at Polymethylmethacrylate Graphite Composite Electrode
  
View Full Text    Download reader
DOI:10.3969/j.issn.1004-4957.年份.月份
KeyWord:nonyl phenol  graphite / PMMA composite electrode  electroanalysis
  
AuthorInstitution
傅清晨,郑少华,纪老三,赖鸣凤,许彩满,黄颖 福建师范大学化学与化工学院
Hits: 2847
Download times: 1261
Abstract:
      In this paper,viscous methylmethacrylate(MMA) was prepolymerized as an hand mixed adhesive for the fabrication of a composite graphite/poly(methylmethacrylate) electrode with high electrochemical performance.Compared with the traditional carbon paste electrode,the electrode fabricated behaves better in mechanical property,stability and reproducibility.The behavior of nonyl phenol(nP)at composite graphite/PMMA electrode was studied by SEM and impedance spectroscopy.Factors influencing the detection processes of nP such as supporting electrolyte,pH value and scan rate were also investigated and optimized.The results indicated that an adsorption-controlled process with 2 electrons and 2 protons transfer occurred at the proposed electrode and the oxidation process for nP was completely irreversible.Under the optimized conditions,a sensitive method of amoperometric measurement was established for the detection of nP at the graphite/PMMA composite electrode.The calibration curve was linear over the nP concentrations in the ranges of 5.0×10-8-7.0×10-7 mol/L(r=0.998)and 9.0 ×10-7-3.0×10-5 mol/L(r=0.996) with a detection limit(S/N=3) of 6.62×10-9mol/L.The proposed method was simple and rapid,and was applied in the determination of nP in real water and soil samples with good recoveries of 94%-103%.
Close