Determination of Sulfaguanidine in Feed Sample by Ni(Ⅳ) Complex-Luminol Chemiluminescence System Coupled with Flow-injection
  
View Full Text    Download reader
DOI:
KeyWord:sulfaguanidine  Ni(Ⅳ) complex  chemiluminescence  luminol  feed  determination
  
AuthorInstitution
SUN Jia-li,LIU Yi,ZHENG Ying-shuai,SONG Ze-an,SHI Hong-mei* 河北医科大学公共卫生学院
Hits: 1995
Download times: 993
Abstract:
      Sulfaguanidine can significantly inhibit the chemiluminescence(CL) intensity of Ni(Ⅳ) complex-luminol system.Based on this,a new method of Ni(Ⅳ) complex-luminol CL system coupled with flow-injection for the determination of sulfaguanidine is developed.Under the optimum conditions,the linear range is 6.0×10-7 mol/L to 6.0×10-6 mol/L to the CL relative intensities with a detection limit(3σ) of 2.4×10-10 mol/L.The RSD(n=11) is 1.4%for 2.4×10-6 mol/L sulfaguanidine.The proposed method was applied in the analysis of sulfaguanidine in swine feed samples with satisfactory results.
Close