Determination of 9 Organophosphate Flame Retardants in Water by Gas Chromatography with Flame Photometry Detection
  
View Full Text    Download reader
DOI:
KeyWord:organophosphate flame retardants  liquid-liquid extraction  gas chromatography with flame photometry detection  water samples
  
AuthorInstitution
PAN Meng,SONG Shu-ling,WANG Yuan,GUO Xiao-chen,RAO Zhu* 国家地质实验测试中心
Hits: 2312
Download times: 1101
Abstract:
      A method for the determination of 9 organophosphate flame retardants(OPFRs) in water by liquid-liquid extraction(LLE) followed by gas chromatography with flame photometry detection was developed.The main analytical conditions,such as water sample preparation and instrument analysis method,were optimized.Liquid-liquid extraction was selected using methylene chloride as extraction solvent,and organic phosphorus Rtx-OPPesticides 2 special column was chosen as analytical column.Under the optimal conditions,a completed base line separation of 9 OPFRs was obtained within 15 min.9 OPFRs in the concentration range of 1.0-500 ng/mL showed good linear relationship(r2≥0.998).The method detection limits(MDLs) ranged from 1.0 ng/L to 2.0 ng/L.At spiked concentrations of 5,100, 300 ng/L,the recoveries of 9 OPFRs were in range of 75.9%-111% with relative standard deviations(RSD) of 3.3%-15%.The method was further applied in the determination of OPFRs in real lake water samples with 9 OPFRs total concentrations of 408-1 532 ng/L.The analytical result showed that the lake water was contaminated by OPFRs.