Detection of Carbofuran by Photocurrent Method Based on TiO2 Molecularly Imprinted Membrane
  
View Full Text    Download reader
DOI:
KeyWord:molecularly imprinted  carbofuran  titanium dioxide  photocurrent  sensor
  
AuthorInstitution
FENG Sha-sha,LIANG Chun-feng,LIANG Shun-chao,WEI Xiao-ping,LI Jian-ping 广西电磁化学功能物质重点实验室,广西高校食品安全与检测重点实验室,桂林理工大学化学与生物工程学院
Hits: 2005
Download times: 918
Abstract:
      A molecularly imprinted photocurrent chemical sensor was fabricated by using TiO2 film for the detection of carbofuran.Effect of doping modification on photocatalytic efficiency of TiO2 was studied.The results showed that the molecularly imprinted membrane of Au doped TiO2 had a good detection effect on the template molecule carbofuran.The experimental conditions such as film thickness and adsorption time were optimized.The experimental results showed that the concentrations of carbofuran had a good liner relationship with photocurrent in the range of 1.00×10-9-2.20×10-7 mol/L under the optimal experimental conditions,and the detection limit(S/N=3) was 1.10×10-10 mol/L.The sensor has demonstrated outstanding features such as excellent sensitivity,selectivity and stability.The determination of carbofuran in water was carried out by using the sensor,and the recoveries were in the range of 98.7%-104.0%.
Close