Preparation of Nano-copper/Graphene-Chitosan Composite Membrane Modified Electrode and Its Direct Determination of Glucose
  
View Full Text    Download reader
DOI:
KeyWord:nano-copper  graphene-chitosan composite membrane  modified electrode  glucose
  
AuthorInstitution
PAN Qian-xiu,CHENG Yuan-zheng,WANG Jiang-yun,ZHANG Feng-lian,DENG Shu-e 潍坊医学院化学教研室
Hits: 1885
Download times: 1230
Abstract:
      The enzyme-free detection of glucose was achieved by using the nano-copper/graphene-chitosan composite matrix modified glass carbon electrode.The microstructure of the composite matrix was characterized by scanning electron microscopy(SEM),FT-IR spectrum and electrochemical impedance spectroscopy(EIS).The electrochemistry behavior of the modified electrode was investigated by cyclic voltammetry and chronoamperometry.The results showed that the modified electrode exhibited a good catalytic oxidation of glucose in 0.1 mol/L NaOH and with a wide linear range of 5.6×10-5-1.2×10-3 mol/L and a low detection limit of 2.3×10-5 mol/L(at signal/noise ratio S/N = 3).The modified electrode could be used to detect the real samples with a good stability and repeatability.
Close