Simultaneous Detection of o-Nitrophenol and p-Nitrophenol Using a β-Cyclodextrin Functionalized Graphene Modified Glassy Carbon Electrode
  
View Full Text    Download reader
DOI:
KeyWord:β-cyclodextrin functionalized graphene  modified electrode  o-nitrophenol  p-nitrophenol  electrochemical behavior  determination
  
AuthorInstitution
DENG Pei-hong*,LIANG Jiao-li,CHEN Dong-er,CAI Fen,LI Qiu-yan,WANG Ting 衡阳师范学院化学与材料科学学院功能金属有机化合物湖南省重点实验室
Hits: 1603
Download times: 626
Abstract:
      A β-cyclodextrin functionalized graphene modified glassy carbon electrode(β-CD-GR/GCE) was prepared by drop-coating method.The electrochemical behaviors of o-nitrophenol(o-NP) and p-nitrophenol(p-NP) on the modified electrode were investigated by cyclic voltammetry and second-order derivative linear sweep voltammetry.The β-CD-GR/GCE showed an excellent electrocatalytical effect on the reduction of p-NP and o-NP.Besides,high resolution capacity to p-NP and o-NP was obtained.Under the optimized conditions,second-order derivative linear sweep voltammetry was utilized for the determination of p-NP and o-NP.There were good linearities for o-NP and p-NP in the concentration ranges of 0.2-10 μmol/L and 0.06-10 μmol/L with detection limits(S/N=3) of 0.08 μmol/L and 0.02 μmol/L,respectively.The proposed method could be applied in the simultaneous determination of o-NP and p-NP in environmental water samples with rapidness and accuracy.
Close