Development and Performance Characterization of a Linear Ion Trap Time of Flight Mass Spectrometer
  
View Full Text    Download reader
DOI:
KeyWord:electrospray ion source  linear ion trap  time of flight mass spectrometer  hybrid mass spectrometer
  
AuthorInstitution
HUANG Xiao,TIAN Qin qin,FAN Rong rong,LI Hui,HUANG Bao,ZOU Xing,ZHU Hui,HUANG Zheng xu 1.昆山禾信质谱技术有限公司;2.广州禾信仪器股份有限公司;3.暨南大学质谱仪器与大气环境研究所
Hits: 2104
Download times: 956
Abstract:
      A high resolution hybrid linear ion trap time of flight mass spectrometer(LIT-TOF MS) was constructed,which is based on the time of flight mass spectrometry,linear ion trap,atmospheric pressure ionization source and other core technologies.The performance of the instrument was characterized with an electric spraying ion source.The resolution of LIT-TOF MS was over 12 000(Reserpine,m/z 609),the mass range was 33-1 922 amu,and the sensitivity was 1 μg/L.It was also demonstrated that MS3 could be readily performed in the LIT.The qualitative and quantitative analyses on amino acids were carried out by the LIT-TOF MS,and results showed that it could be used in the real time,on line and high throughput determination of contents and variations of amino acids in biological complex samples.Besides,the instrument,combined with a variety of atmospheric pressure ionization techniques,could be applicable for determination in fields of medicine,environment and foods,etc.