Detection of 73 Glucocorticoids in Cosmetics by Ultra Performance Liquid Chromatography-Tandem Mass Spectrometry with Pass-through Column Purification
  
View Full Text    Download reader
DOI:
KeyWord:cosmetics  glucocorticoids  ultra performance liquid chromatography-tandem mass spectrometry(UPLC-MS/MS)
  
AuthorInstitution
LIU Hong,YANG Piao piao,LI Li-xia 1.Hubei Institute for Drug Control;2.Hubei Engineering Research Center for Drug Quality Control
Hits: 1643
Download times: 977
Abstract:
      An ultra performance liquid chromatography-tandem mass spectrometric(UPLC-MS/MS) method was established for the determination of 73 glucocorticoids in cosmetics.Samples were dispersed with saturated sodium chloride and ultrasonically extracted with acetonitrile,then purified with a PRiME HLB SPE column(60 mg/3 mL) and separated on a Waters Cortecs C18 column(150 mm×2.1 mm,2.7 μm) by gradient elution using 0.1% acetic acid solution and acetonitrile as mobile phases.The analysis was performed with electrospray ion source in positive ion mode under multi-reaction monitoring(MRM) mode,and the results were quantified by external standard method.Results showed that there were good linear relationships for 73 glucocorticoids in cosmetics in their respective concentration ranges with correlation coefficients(r) higher than 0.99.The limits of quantitation(LOQ) were in the range of 0.1-0.3 μg/g.The average recoveries for 73 glucocorticoids at three spiked levels of 1LOQ,2LOQ and 5LOQ ranged from 73.3% to 119%,with relative standard deviations(RSD) of 0.40%-18%.The method was rapid and accurate,and could be used as a screening approach for 73 glucocorticoids in cosmetics.