Detection of 7 Volatile Sulfur Compounds in Plastic Track Surface by Ultrasonic Extraction-HS-SPME/GC-MS Technique
  
View Full Text    Download reader
DOI:
KeyWord:headspace solid phase microextraction(HS-SPME)  gas chromatography-mass spectrometry(GC-MS)  plastic track surface  volatile sulfur compounds(VSCs)
  
AuthorInstitution
YE Yuan-jian,ZHENG Yu,CAI Jin-an,LIU Hai-yan,LU Jia-jun,LI Xiao-zeng,CHEN Wei-li Guangzhou Quality Supervision and Testing Institute,Guangzhou ,China
Hits: 1021
Download times: 672
Abstract:
      A gas chromatography-mass spectrometry(GC-MS)coupled with ultrasonic extraction and headspace solid phase microextraction(HS-SPME) was established for the rapid analysis of seven volatile sulfur compounds(VSCs),including dimethyl sulfide(DMS),carbon disulfide(CS2),methyl ethyl sulfide(MES),diethyl sulfide(DES),dimethyl disulfide(DMDS),diethyl disulfide(DEDS),dimethyl trisulfide(DMTS) in plastic track surface.The HS-SPME extraction conditions,ultrasonic extraction conditions and GC-MS parameters were optimized.Under the optimized conditions,7 VSCs had good linearities in the range of 5-500 ng/g for DES and DMDS,and 10-500 ng/g for other five VSCs,with correlation coefficients(r) of 0.984 7-0.998 9.The limits of detection(LODs,S/N = 3) were in range of 0.4-1.4 ng/g,and the limits of quantitation(LOQs,S/N = 10) were 1.3-4.7 ng/g.The average recoveries at low,medium and high spiked levels ranged from 89.6% to 104%,with relative standard deviations(RSDs,n = 6) of 3.5%-12%.The method integrates separation,enrichment and injection into one step,and is suitable for the routine detection of VSCs in the synthetic surfaces.