Optimization of Measuring Methods for Methane Clumped Isotopologues 12CH2D2
  
View Full Text    Download reader
DOI:
KeyWord:methane  clumped isotope  high-resolution gas stable isotope ratio mass spectrometer(HR-IRMS)  precision
  
AuthorInstitution
LU Li,WANG Xin-chu,LI Si-liang,PANG Zhi-yong 1. Institute of Surface-Earth System Science,School of Earth System Science,Tianjin University, Tianjin ,China; 2. Haihe Laboratory of Sustainable Chemical Transformations, Tianjin University,Tianjin ,China
Hits: 1205
Download times: 895
Abstract:
      Recently developed methane clumped isotope techniques provide fresh and novel insights into methane production and transformation,which have been unobtainable through other techniques such as traditional stable isotope.Methane clumped isotope techniques have the advantages of high resolution and low detection limit.13CH3D and 12CH2D2 have their unique chemical and physical properties recording information on methane isotopic equilibrium states.The measuring methods for paired clumped isotopologues have just been established in China,and the precision of δ12CH2D2 need to be further improved.Based on the first high-resolution gas stable isotope mass spectrometer(HR-IRMS) in China,the measuring method for 12CH2D2 was optimized by balancing the resolution and sensitivity of the instrument.The [12CH2D2+] intensity was effectively improved.The analysis precision was successfully improved to 1.29‰ by using H2O+ peak simulation calculation for the contribution of adjacent 13CH3D and 13CH5 peaks to the total background.